‍Oximeter 5 Click is a compact add-on board suitable for measuring blood oxygen saturation. This board features the MAX30102, integrated pulse oximetry, and heart-rate monitor module from Analog Devices. The MAX30102 includes internal LEDs, photodetectors, optical elements, and low-noise electronics with ambient light rejection. It operates on a single 1.8V power supply acquired from both mikroBUS™ power rails for the internal LEDs, communicating through a standard I2C compatible interface. The MAX30102 can be shut down through software with zero standby current, allowing the power rails to remain powered at all times.

click Product page


Click library

  • Author : Nenad Filipovic
  • Date : Nov 2021.
  • Type : I2C type

Software Support

We provide a library for the Oximeter 5 Click as well as a demo application (example), developed using MikroElektronika compilers. The demo can run on all the main MikroElektronika development boards.

Package can be downloaded/installed directly from NECTO Studio Package Manager(recommended way), downloaded from our LibStock™ or found on Mikroe github account.

Library Description

‍This library contains API for Oximeter 5 Click driver.

Standard key functions :

Example key functions :

  • oximeter5_read_sensor_data Oximeter 5 get sensor data function.
    err_t oximeter5_read_sensor_data ( oximeter5_t *ctx, uint32_t *ir, uint32_t *red );
    err_t oximeter5_read_sensor_data(oximeter5_t *ctx, uint32_t *ir, uint32_t *red)
    Oximeter 5 get sensor data function.
  • oximeter5_get_oxygen_saturation Oximeter 5 get oxygen saturation function.
    err_t oximeter5_get_oxygen_saturation ( uint32_t *pun_ir_buffer, int32_t n_ir_buffer_length, uint32_t *pun_red_buffer, uint8_t *pn_spo2 );
    err_t oximeter5_get_oxygen_saturation(uint32_t *pun_ir_buffer, int32_t n_ir_buffer_length, uint32_t *pun_red_buffer, uint8_t *pn_spo2)
    Oximeter 5 get oxygen saturation function.
  • oximeter5_read_temperature Oximeter 5 read temperature function.
    err_t oximeter5_read_temperature ( oximeter5_t *ctx, float *temperature );
    err_t oximeter5_read_temperature(oximeter5_t *ctx, float *temperature)
    Oximeter 5 read temperature function.

Example Description

‍This library contains API for Oximeter 5 Click driver. The demo application reads and calculate SpO2 oxygen saturation data.

The demo application is composed of two sections :

Application Init

‍Initializes I2C driver and log UART. After driver initialization the app set driver interface setup and default settings, buffer length of 100 stores 4 seconds of samples running at 25sps read the first 100 samples, and determine the signal range.

void application_init ( void )
{
log_cfg_t log_cfg;
oximeter5_cfg_t oximeter5_cfg;
LOG_MAP_USB_UART( log_cfg );
log_init( &logger, &log_cfg );
log_info( &logger, " Application Init " );
// Click initialization.
oximeter5_cfg_setup( &oximeter5_cfg );
OXIMETER5_MAP_MIKROBUS( oximeter5_cfg, MIKROBUS_1 );
if ( I2C_MASTER_ERROR == oximeter5_init( &oximeter5, &oximeter5_cfg ) )
{
log_error( &logger, " Communication init." );
for ( ; ; );
}
Delay_ms ( 100 );
if ( OXIMETER5_ERROR == oximeter5_default_cfg ( &oximeter5 ) )
{
log_error( &logger, " Default configuration." );
for ( ; ; );
}
Delay_ms ( 100 );
un_brightness = 0;
un_min = 0x3FFFF;
un_max = 0;
for ( uint8_t n_cnt = 0; n_cnt < 100; n_cnt++ )
{
oximeter5_read_sensor_data( &oximeter5, &aun_red_buffer[ n_cnt ], &aun_ir_buffer[ n_cnt ] );
if ( un_min > aun_red_buffer[ n_cnt ] )
{
un_min = aun_red_buffer[ n_cnt ];
}
if ( un_max < aun_red_buffer[ n_cnt ] )
{
un_max = aun_red_buffer[ n_cnt ];
}
}
oximeter5_get_oxygen_saturation( &aun_ir_buffer[ 0 ], 100, &aun_red_buffer[ 0 ], &n_spo2 );
log_info( &logger, " Application Task " );
Delay_ms ( 100 );
}
#define OXIMETER5_MAP_MIKROBUS(cfg, mikrobus)
MikroBUS pin mapping.
Definition oximeter5.h:217
#define OXIMETER5_INTERRUPT_ACTIVE
Definition oximeter5.h:192
uint8_t oximeter5_check_interrupt(oximeter5_t *ctx)
Oximeter 5 check interrupt function.
void application_init(void)
Definition main.c:45
@ OXIMETER5_ERROR
Definition oximeter5.h:265

Application Task

‍This is an example that demonstrates the use of the Oximeter 5 Click board™. In this example, display the IR and RED ADC data, and the SpO2 oxygen saturation data [ 0% - 100% ]. Results are being sent to the Usart Terminal where you can track their changes.

void application_task ( void )
{
for ( uint8_t n_cnt = 25; n_cnt < 100; n_cnt++ )
{
aun_red_buffer[ n_cnt - 25 ] = aun_red_buffer[ n_cnt ];
aun_ir_buffer[ n_cnt - 25 ] = aun_ir_buffer[ n_cnt ];
if ( un_min > aun_red_buffer[ n_cnt ] )
{
un_min = aun_red_buffer[ n_cnt ];
}
if ( un_max < aun_red_buffer[ n_cnt ] )
{
un_max=aun_red_buffer[n_cnt];
}
}
for ( uint8_t n_cnt = 75; n_cnt < 100; n_cnt++ )
{
un_prev_data = aun_red_buffer[ n_cnt - 1 ];
oximeter5_read_sensor_data( &oximeter5, &aun_red_buffer[ n_cnt ], &aun_ir_buffer[ n_cnt ] );
if ( aun_red_buffer[ n_cnt ] > un_prev_data )
{
f_temp = aun_red_buffer[ n_cnt ]-un_prev_data;
f_temp /= ( un_max - un_min );
f_temp *= MAX_BRIGHTNESS;
f_temp = un_brightness - f_temp;
if ( f_temp < 0 )
{
un_brightness = 0;
}
else
{
un_brightness = ( uint32_t ) f_temp;
}
}
else
{
f_temp = un_prev_data - aun_red_buffer[ n_cnt ];
f_temp /= ( un_max - un_min );
f_temp *= MAX_BRIGHTNESS;
un_brightness += ( uint32_t ) f_temp;
if ( un_brightness > MAX_BRIGHTNESS )
{
un_brightness = MAX_BRIGHTNESS;
}
}
if ( ( OXIMETER5_OK == oximeter5_get_oxygen_saturation( &aun_ir_buffer[ 0 ], 100, &aun_red_buffer[ 0 ], &n_spo2 ) ) )
{
if ( aun_ir_buffer[n_cnt] > 10000 )
{
log_printf( &logger, "\tIR : %lu \r\n", aun_ir_buffer[ n_cnt ] );
log_printf( &logger, "\tRED : %lu \r\n", aun_red_buffer[ n_cnt ] );
log_printf( &logger, "- - - - - - - - - - - - - - -\r\n" );
log_printf( &logger, "\tSPO2 : %d %%\r\n", ( uint16_t ) n_spo2 );
log_printf( &logger, "-----------------------------\r\n" );
Delay_ms ( 100 );
}
else
{
Delay_ms ( 10 );
}
}
}
}
#define MAX_BRIGHTNESS
Definition oximeter5.h:187
void application_task(void)
Definition main.c:106
@ OXIMETER5_OK
Definition oximeter5.h:264

Note

‍A measurement time of at least 10 seconds is required for the SpO2 oxygen saturation data to be valid.